ridm@nrct.go.th   ระบบคลังข้อมูลงานวิจัยไทย   รายการโปรดที่คุณเลือกไว้

The reliability of the i-STAT clinical portable analyser

หน่วยงาน Central Queensland University, Australia

รายละเอียด

ชื่อเรื่อง : The reliability of the i-STAT clinical portable analyser
นักวิจัย : Dascombe, Benjamin James. , Reaburn, Peter. , Sirotic, Anna. , Coutts, Aaron James.
คำค้น : Blood , Healthy Communities , 730299 Public health not elsewhere classified , 321401 Exercise Physiology , Exercise. , Reliability -- Blood analytes -- Field research -- Blood lactate
หน่วยงาน : Central Queensland University, Australia
ผู้ร่วมงาน : -
ปีพิมพ์ : 2550
อ้างอิง : http://hdl.cqu.edu.au/10018/6226 , , cqu:2537
ที่มา : Dascombe, B, Reaburn, P, Sirotic, A & Coutts, A 2007, 'The reliability of the i-STAT clinical portable analyser', Journal of Science and Medicine in Sport, vol. 10, no. 3, pp. 135-140, http://dx.doi.org/10.1016/j.jsams.2006.05.023
ความเชี่ยวชาญ : -
ความสัมพันธ์ : Journal of Science and Medicine in Sport : Amsterdam, Netherlands. : Elsevier. Vol 10, issue 3 (2007), p. 135-140 6 pages Refereed 1440-2440 , aCQUIRe [electronic resource] : Central Queensland University Institutional Repository.
ขอบเขตของเนื้อหา : -
บทคัดย่อ/คำอธิบาย :

The purpose of this investigation was to assess the reliability of the i-STAT clinical portable analyser and CG(4)(+) cartridge measures of blood pH, partial pressures of O(2) (pO(2)) and CO(2) (pCO(2)), bicarbonate ([HCO(3)(-)]), base excess (BE), total carbon dioxide (TCO(2)), oxygen saturation (sO(2)) and blood lactate ([BLa(-)]) at various exercise intensities. A comparison between [BLa(-)] measured by the i-STAT and the Accusport lactate analysers during an intermittent treadmill run was also undertaken. The technical error of measurement (TEM%) at rest, at moderate (200W) and maximal exercise (V O(2)max) was acceptable (<15%) for all parameters. The intra-class correlation coefficients for each analyte ranged from weak-to-strong across resting (ICC=0.19-0.96) and moderate (ICC=0.30-0.96) exercise intensities. The ICC for all analytes were observed to be strong following maximal intensity exercise (ICC=0.77-0.95). The comparison of the [BLa(-)] measures between the i-STAT and Accusport showed that the difference between measures was acceptable at both low (<4mmolL(-1)) (-0.39+/-0.27mmolL(-1)), moderate to high concentrations (>4mmolL(-1)) (0.58+/-1.22mmolL(-1)), and across all [BLa(-)] data (0.36+/-1.13mmolL(-1)). In conclusion, the i-STAT clinical analyser and CG(4)(+) cartridge provides reliable measures of a number of blood parameters across exercise intensities. The [BLa(-)] measures from the i-STAT analyser are consistent with that of the Accusport lactate analyser.

บรรณานุกรม :
Dascombe, Benjamin James. , Reaburn, Peter. , Sirotic, Anna. , Coutts, Aaron James. . (2550). The reliability of the i-STAT clinical portable analyser.
    กรุงเทพมหานคร : Central Queensland University, Australia.
Dascombe, Benjamin James. , Reaburn, Peter. , Sirotic, Anna. , Coutts, Aaron James. . 2550. "The reliability of the i-STAT clinical portable analyser".
    กรุงเทพมหานคร : Central Queensland University, Australia.
Dascombe, Benjamin James. , Reaburn, Peter. , Sirotic, Anna. , Coutts, Aaron James. . "The reliability of the i-STAT clinical portable analyser."
    กรุงเทพมหานคร : Central Queensland University, Australia, 2550. Print.
Dascombe, Benjamin James. , Reaburn, Peter. , Sirotic, Anna. , Coutts, Aaron James. . The reliability of the i-STAT clinical portable analyser. กรุงเทพมหานคร : Central Queensland University, Australia; 2550.