| ชื่อเรื่อง | : | Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene |
| นักวิจัย | : | Kilgour, Alastair Scott |
| คำค้น | : | total hip arthroplasty , Polymer , Scanning electron , wear |
| หน่วยงาน | : | Edinburgh Research Archive, United Kingdom |
| ผู้ร่วมงาน | : | Elfick, Alistair , Royal Society of Edinburgh |
| ปีพิมพ์ | : | 2553 |
| อ้างอิง | : | http://hdl.handle.net/1842/4628 |
| ที่มา | : | - |
| ความเชี่ยวชาญ | : | - |
| ความสัมพันธ์ | : | Atwood, S.A., Kilgour, A., Hoang, M., Furmanski, J., Pruitt, L.A., Lamellar orientation of plastically strained ultra-high molecular weight polyethylene, (Poster) presented at: Second International Conference on Mechanics of Biomaterials and Tissues, Lihue, Kaua'l Hawaii, USA, Dec. 2007. , Kilgour, A. and Elfick, A.P., Design and test of novel tribometer: investigation into wear performance of materials used in total hip replacements. Tribology - Materials, Surfaces & Interfaces, 2008. 2: p. 161-168. , Kilgour, A. and Elfick, A.P., Investigation of Polyethylene structure and its bearing on wear of joint replacements. (Abstract). Oral presentation In: World Biomaterials Congress. Amsterdam, 2008. , Kilgour, A. and Elfick, A.P., Influence of Crosslinked Polyethylene Structure on Wear of Joint Replacements. (Abstract). Oral presentation In: 35th Leeds-Lyon Symposium on Tribology, Leeds, 2008. , Kilgour, A. and Elfick, A.P., Influence of crosslinked polyethylene structure on wear of joint replacements. Tribology International, 2009. 42(11-12): p. 1582-1594. |
| ขอบเขตของเนื้อหา | : | - |
| บทคัดย่อ/คำอธิบาย | : | Total hip arthroplasty commonly involves a hard metallic/ceramic femoral ball component articulating against an acetabular ultra-high molecular weight polyethylene (UHMWPE) counter-bearing. A novel six-station, wear tribometer, featuring programmable load and motion, was designed to further the investigation into wear, sub-surface plasticity and debris generation of UHMWPE. This thesis describes the pin-on-plate device, its validation and subsequent use to assess unirradiated (-PE) and gamma-irradiated highly crosslinked (+PE) UHMWPE wear behaviour. With the emphasis on dynamic loading and a closer gait matched open wear path, the tribometer improves on the clinical relevance of pin-on-plate testing. There is a requirement for this type of machine in order to investigate the directional dependence of wear and debris generation of UHMWPE more accurately, where “simplified” tribometers (adequate for constant load/constant velocity and constant load/sinusoidal velocity work) are not capable or suitable. For the first time in orthopaedic pin-on-plate studies, tests were conducted using an advanced dynamic load synchronised to a more physiologically accurate elliptical motion path. To validate the machine, three orthopaedic polymers of clinical relevance; Polytetrafluroethylene, Polyacetal, and UHMWPE were subjected to linear-reciprocating (LR) and novel elliptical motion paths under a Paul-type load profile. All three polymers showed higher wear factors under elliptical motion, by up to 2 orders of magnitude, agreeing well with explanted values. The UHMWPE elliptical wear factor was comparable to that reported for clinical, where kelliptical = 1.56 x 10-6 mm3/Nm. In the crosslinked study, the mean steady state wear of -PE and +PE groups under linear reciprocating motion was not significantly different. However, under elliptical motion, crosslinking reduced UHMWPE wear by up to 92% when compared to the unirradiated group. In –PE pins worn under LR motion and in +PE pins subjected to both motion paths a sub-surface damage zone with reduced crystallinity and increased strain was measured using Raman spectroscopy. This was attributed to large strain accumulation in the slower wearing surfaces providing a mechanism for de-crystallisation. The discovery of such a near-surface layer is in good agreement with critical strain wear models. In disagreement, however, we found the sliding induced layer to extend to greater depths than previous reported. |
| บรรณานุกรม | : |
Kilgour, Alastair Scott . (2553). Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene.
กรุงเทพมหานคร : Edinburgh Research Archive, United Kingdom . Kilgour, Alastair Scott . 2553. "Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene".
กรุงเทพมหานคร : Edinburgh Research Archive, United Kingdom . Kilgour, Alastair Scott . "Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene."
กรุงเทพมหานคร : Edinburgh Research Archive, United Kingdom , 2553. Print. Kilgour, Alastair Scott . Tribometer with programmable motion and load to investigate the influence of molecular structure on wear of orthopaedic polyethylene. กรุงเทพมหานคร : Edinburgh Research Archive, United Kingdom ; 2553.
|